Catalogue values for Spectrum Analyzers and Vector Network Analyzers
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
FEATURES | Description of the itemuu | Spectrum analyzers |
GENERAL FEATURES COMMON TO SPECTRUM AND VECTOR ANALYSER | Markers | Normal, delta, 12 markers |
Trigger | Free run, video, External | |
Display | Internal TFT/LCD Color Display with provision for six traces | |
Interface | USB, LAN | |
Software for transfer of screen shot/measurement data to PC | Yes | |
SPECTRUM ANALYZER FEATURES | Frequency Range | 10 Hz to 26.5 GHz |
Aging rate of spectrum analyzer frequency: Better than 1x10-6 /Year | Yes | |
Frequency counter Resolution(minimum) (Hz) | 0.001Hz | |
Frequency Span | 10 Hz to 26.5 GHz | |
Spectral purity:SSB Phase Noise @10KHz offset shall be less than -90dBc/Hz at 1GHz | SSB Phase Noise @ 10KHz offset , at 1GHz shall be less than -100dBc/Hz | |
Spectral purity:Harmonics @ -40dBm level <=60 dBc | yes | |
Sweep time @span >or= 1 KHz | 1 mS to 4000 S | |
Sweep time @span = 0 Hz | 1microS - 1000sec | |
Resolution Bandwidth (-3dB) | 1Hz to 10MHz in steps | |
Video Bandwidth | 1Hz to 10MHz in steps. | |
Amplitude : Max input level for protection (dBm) | +30 | |
Amplitude : Max input level for measurement (dBm) | 20 | |
Amplitude :level measurement uncertainity (dB) | <=1, <=±1 | |
Average noise level displayed(@300HzRBW,10Hz VBW) | <=-140dBm for 10MHz to 3 GHz | |
Average noise level displayed(@100HzRBW,1Hz VBW) | <= -140dBm for 13GHz to 26 GHz, <= -140dBm For 3GHz to 13GHz | |
Average noise level displayed(@300HzRBW,1Hz VBW) | <=-110dBm for 10MHz to 3 GHz, <= -125dBm for 3GHz to 13 GHz | |
Measurement features | C/N ratio,AM%,time domain power measurement,channel power measurement,Phase noise measurement,TOI(THIRD ORDER INTERCEPT),OBW(occupied bandwidth) | |
VECTOR NETWORK ANALYZER FEATURES | Cable Loss (dB) | NA |
Calibration with Precision calibration components-open, short, load should be included to achieve directivity >35 dB hint: NA for items designated as spectrum analyzers | NA | |
RF output port | NA | |
Nominal port power output: Hi Power/Lo-Power selectable | NA | |
OPERATING CONDITIONS | Operating Temperature | 0 to 50 degree C & 80% RH @ 40 degree C |
Operating Humidity | 90% @ 40deg. C | |
POWER REQUIREMENT | Power requirement | 230 +- 10 VAC, 50Hz |
WARRANTY | Warranty | 3, 4, 5 Or higher |
CERTIFICATION/TEST REPORT DETAILS | The equipment supplied with calibration certificate | yes |
conformity to Dry Heat for 16 hrs at a specified maximum temp in accordance with IS:9000/part-3/section-5/1977 latest | no | |
conformity to Cold Test for 4 hrs at a specified minimum temp in accordance with IS:9000/part-2/section-4/1977 latest | no | |
conformity to Damp Heat (Cyclic)Test for 2 Cycles of 24 h each at a temp of 40 degree C & 95% RH | no | |
Availability of Test Reports from Central Govt/NABL/ILAC approved accredited lab to prove conformity to the specification | Yes | |