Catalogue values for Anaesthesia Workstation (V2)

Category Specification Bid Requirement (Allowed Values)
GENERAL Product Description Anesthesia Workstation
Components of Anesthesia workstation Anesthesia machine , Inbuilt ventilator , vaporisor , Gas delivery system , Multi para monitor
Gas Delivery System Availability of Pin index yokes for gases Oxygen and Nitrous oxide
Number of Yokes, Oxygen 1
Number of Yokes, N2O 1
NIST /DISS Inlet pipeline with Hoses for O2, N2O and Air Yes
Availability of flow meters for O2, N2O and air Yes
Type of Flowmeter Rotameter
Mode of gas mixing Pneumatic
Availability of pressure gauges for cylinder and pipelines in anesthesia workstation Yes
Availability of audible and visual oxygen failure alarms Yes
Minimum emergency oxygen flow (in liter per minute) 35 - 70
Type of hypoxic guard with automatic cutoff of N2O Mechanical
Minimum 25% O2 across all Oxygen - Nitrous Oxide Mixture Yes
Auxillary Oxygen Flowmeter Yes
Auxillary Common Gas outlet for use in open circuit Yes
Vaporizer Number of quick mount type vaporizer for easy interchangeability at a time 2
Number of selectatec back bar 1
Facility of interlock for vaporizer Yes
Minimum agent capacity for free volatile anesthetic agent (in ml) 225 ml
Sensor connection shall be internal to prevent disconnection Yes
Vaporizer compatible to gases Sevoflurane, Isoflurane, Desflurane, Halothane
The vaporizer design should be maintenance free Yes
Vaporizer calibration requirement After every 5 years
Should have a drug specific Filling arrangement Yes
Breathing System Breathing system should have fresh gas de-coupled / compensation , closed circle absorber system Yes
Range of pressure relief valve (in mbar) of breathing system 5 to 75
Number of steps in change over from Mechanical / Spontaneous to bag ventilation Single step
Number of reusable autoclavable closed silicone circuit supplied 2 Nos for adult
Number of Disposable closed circuits - Adult NA
Number of Disposable closed circuits - Pediatric NA
Number of Disposable closed circuits - Neonatal NA
Number of bains circuit (adult) supplied with workstation 5
Number of Jackson Rees circuit for pediatric patients supplied with workstation 5
Volume capacity of breathing system (in liters) 1.5 or more
Reusable flow sensors covered in warranty No
Dual flow sensor at inspiratory and expiratory port No
Breathing system should have leak and compliance test Yes
Breathing system should be autoclavable Yes
Facility of CO2 Bypass No
Anesthesia Ventilator Available modes of operating ventilator Manual/spontaneous, Volume controlled, Pressure controlled, SIMV/PCV, SIMV/VCV, CMV
Ventilator should be capable to pressure support Yes
Ventilator should be capable to dual mode of ventilation Yes
Integrated ventilator not require bellows to be change for adults and infants Yes
Tidal volume of ventilator (in ml) 10 to 1400 ml
PEEP of ventilator in mbar 0 to 20 mbar
Breathing frequency in BPM 4 to 60 BPM
Inspiratory pause percentage of Ti 0 - 50%
Screen size of Ventilator monitor in inches 6 to 10, 10.1 to 15, 15 .1 and more Or higher
Integrated ventilator shall be able to provide facility to monitor FiO2, dual agent, CO2, Automatic identification of agents, Minimal Alveolar Concentration (MAC) Value Yes, No
Integrated ventilator shall be able to provide facility to monitor expired O2 (EtO2) No
Flow Trigger Sensitivity Adult, Pediatric, Neonatal
Monitor Type of monitor LCD/LED/TFT Yes
Size of monitor in inches 12 to 15, 15.1 to 18, 18.1 and more Or higher
Monitor should not require lengthy start up procedure or calibration for display of parameters Yes
The graphical and numerical trend with split screen duration 24 hours
Patient multipara monitor shall be able to display parameters HR, SpO2
Capnography to be displayed Yes
IBP1 and IBP2 to be displayed Yes
CO (cardiac output) to be displayed No
Anesthesia Gas Monitoring (AGM) No, Yes
BIS/ENTROPY to be displayed No
Neuro Muscular Transmission(NMT) to be displayed No
ECG monitoring with standard lead configuration, I,II,III, avL, avF&V filters so that it is protected against defibrillator, electro surgery potentials, arrhythmia analysis and ST segment analysis with trends Yes
NIBP should be determine with oscillometric method, should read systolic, diastolic and mean arterial pressure and pulse rate Yes
Pulse oximetery with sensors suitable for adults, pediatric & neonate Yes
ETCO2 measurement should be by infra –red absorption technique,measuring mode should be side streamor main stream and response time should be less than 3 seconds Yes
Monitor should be able to measure and display all parameters in single screen Yes
Monitor should be provided from the same OEM Yes
Alarms Alarms Tidal Volume, Minute Volume, Airway Pressure Low and High, Inspiratory Oxygen concentration, Audio power supply fail alarm, Fail to cycle warning (apnea).
Alarms shall be adjustable to high / low limits with audio and visual alarms Yes
Power supply Length of main cable in meter with power socket of standard indian make to be provided ≥5
Inbuilt battery backup for monitor in Mins 120, 180, 240, 300, 360 Or higher
Power supply 220-240V AC, 50 Hz single phase
Additional Accessories Transparent re-usable silicon face mask size 0,1, 2, 3, 4 & 5 2 each
Breathing bag capacity - 2 Each 1 Litre
3/5 ECG lead 2
Nasopharyngeal/rectal and skin Temperature probes should be Provided 2 Nos Adult
NIBP Reusable Large adult, Reusable small adult cuffs, Reusable Pediatric cuff 2 each
ETCO2 sample line 10
Pressure regulator with 3 meter high Pressure tubing(1 for oxygen and 1 for Nitrous oxide) Yes
Provision for laryngscopes Not Provided
Number of Reusable patient circuits Not provided
SpO? sensor with cable provided 2 Nos Adult
Reusable IBP interface cables Not provided
Disposable IBP transducers Not provided
BIS sensors Not provided
Accessories for wall mounting NA (If trolley mounting)
ADDITIONAL PARAMETERS Type of mounting for the Anesthesia workstation Trolley Mounting
Material used for Anesthesia workstation Epoxy powder painted steel
Mode of integrated scavenging system with pressure relief valve Passive
Minimum Battery backup for anesthesia workstation 1 hour
Facility of manual ventilation in case of electricity and battery failure Yes
Facility of gas and agent delivery in case of electricity and battery failure Yes
All the components likes anesthesia machine, integrated ventilator, vaporizer from the same manufacturer Yes
The workstation should have one or more drawers to keep accessories, good mobility, anti-static caster wheels with locking facility & conveniently placed handles for easy movement of the machine Yes
All software updates should be provided free of cost during Warranty period Yes
The anesthesia machine can be used safely up to 1000 gauss line while being used with MRI No
Pressure relief 5 to 70 cm H20
CERTIFICATIONS Compliance to Medical Device Rules (MDR) 2017 as amended till date Yes
Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date Yes
Certification for manufacturing unit ISO:13485 (Latest)
Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date Yes
Electrical Protection certification IEC-60601-1-1 or equivalent BIS
Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement Yes
WARRANTY Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 3, 4, 5 Or higher