Catalogue values for Gas Chromatography Mass Spectrometry (GC - MS)
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
GAS CHROMATOGRAPH (GC) CONFIGURATION | "GC Detectors inclusive in the scope of supply | Flame Ionization Detector (FID), Thermal Conductivity Detector (TCD or Hot Wire Detector), Electron Capture Detector (ECD), Nitrogen–Phosphorus Detector (NPD) / Flame Thermionic Detector (FTD), Flame Photometric Detector (FPD), Photo Ionization Detector (PID), Pulsed Flame Photometric Detector (PFPD) |
"Number of FID in the System - inclusive in the scope of Supply | 0, 1, 2 | |
"Number of TCD in the System - inclusive in the scope of Supply | 0, 1, 2 | |
"Number of ECD in the System - inclusive in the scope of Supply | 0, 1, 2 | |
"Number of NPD/FTD in the System - inclusive in the scope of Supply | 0, 1, 2 | |
"Number of FPD in the System - inclusive in the scope of Supply | 0, 1, 2 | |
"Number of PID in the System - inclusive in the scope of Supply | 0, 1, 2 | |
"Number of PFPD in the System - inclusive in the scope of Supply | 0, 1, 2 | |
Number of Inlets GC can Simultaneously Support | 2 | |
Number of Detectors (Ports) GC can Simultaneously Support | 2 | |
Number of Capillary Columns provided in GC - inclusive in the scope of Supply | 2 | |
Type of Capillary Column | Polar, Non-Polar | |
"GC Capillary Column Length (Hint:-Select '0' if Not Applicable)" | 30.0 | |
GC Packed Column Length | 0, 3, 6, 9, 12 | |
PERFORMANCE | GC Detector Minimum Detection Level (MDL) | < 1.4 pg C/s for FID, < 1.8 pg C/s for FID, < 3.0 pg C/s for FID < 400 pg/ml for TCD, < = 1 microgram/ml for TCD, < = 10 microgram/ml for TCD, < 4.5 fg /sec for ECD, < 5.5 fg /sec for ECD, < 0.1pg N /sec for NPD/FTD, < 0.03 pg P /sec for NPD/FTD, < 45 fg P /sec for FPD, < 100 fg P /sec for FPD, <2.5 pg S /sec for FPD, <5 pg S /sec for FPD |
COLUMN OVEN (GAS CHROMATOGRAPH - GC) | Maximum Operating Temperature (GC Oven) | upto 400 deg. C, upto 450 deg. C (2) Or higher |
Number of Oven Ramps / Plateaus for Temperature programming | 6/7 or more, 9/10 or more, 20/21 or more | |
PNEUMATIC CONTROL | Number of Injection Ports | 2 |
Maximum Injection Port heating Temperature | upto 400 deg. C (1), upto 450 deg. C (2) Or higher | |
Pressure Range | upto 120 psi (2) Or higher | |
Pressure set point Resolution (±)(psi) | 0.001, 0.01, 0.1 Or lower | |
Type of Sample Injection Port | split / Splitless | |
Split Ratio | 6000:1, 7500:1 | |
AUTO INJECTOR / AUTO SAMPLER | Compatible Liquid Auto Sampler - inclusive in the scope of supply | Yes |
Capacity of the Liquid Auto Sampler (Number of Vials) | 100, 150 Or higher | |
Number of Injection Ports in the Liquid Auto Sampler | 1, 2 | |
Number of Channels in the Liquid Auto Sampler | 0, 2 | |
Liquid Auto Sampler Features | Fully automated Liquid Auto Sampler, Injects with variable speed & varying syringe sizes, Operates through Software (Software inclusive in the scope of supply) | |
HEAD SPACE SAMPLER | Compatible Head Space Sampler- inclusive in the scope of supply | Yes, No |
Capacity of the Head Space Sampler Carousel (number of Vials) | 12 or better, NA | |
Head Space Sampler Temperature Range | Upto 200 deg. C or better, NA | |
PYROLYZER SYSTEM | Compatible Pyrolyzer System for pyrolysis of solid and liquid samples- inclusive in the scope of supply | Yes, No |
MASS SPECTROPHOTOMETER (MS) SYSTEM | Ionization modes | Electron Ionization (EI) |
Detector Type | Electron Multiplier | |
Resolution | 1 amu | |
Scan Rate | upto 20000 amu/sec or better | |
Operating modes | Full Scan Mode, Selective Ion Monitoring (SIM) Mode, Both Full Scan & SIM Modes | |
PERFORMANCE OF THE MASS SPECTROPHOTOMETER (MS) SYSTEM | Sensitivity (S/N Ratio) in full scan | 1500:1 with 1 pg/microliter OFN, 800:1 (EI) or better OFN, 700:1 (EI) or better OFN, 250:1 (EI) or better OFN, 300:1 (PCI) or better BZP, 2000:1 (NCI) or better OFN |
SALIENT FEATURES OF GC-MS | "Salient Featutes of GC-MS (Select applicable features Only)" | Automatic leak testing/detection of Pneumatics, Direct setting of split flow rates and ratios, Flow sensor for control & storage of split ratio, Split and Septum Purge gases, Able to direct inject onto packed and wide-bore capillary columns, Oven power turns-off automatically when the lid/door is opened, Over Heat Protection, Display Unit with built-in Key Board, Controlled by External PC System, Programmable ion source temperature, System includes variety of auto tune algorithms to tune the instrument for maximum sensitivity or for specific target compounds, EI source is inert to active compounds and programmable, EI Source have a dual filament design with automatic selection of other filament if one fails during analysis; |
LIBERARIES | Libraries for GC-MS | NIST Library for MS - Licensed version - 2017 or latest, Wiley Library |
SOFTWARE | "Software Features (Select Applicable Features only)" | Compatible with Windows 7 or higher version operating system, Seamless integration & control of all the GC parameters, Interface with Multiple GCs through the comprehensive real-time monitoring and diagnostic software, Real time notification via advisories and indicators, Counters, Chromatographic attributes, Instrument diagnostics, Access to maintenance information, Access to maintenance, run and event logs, Retention Time - Locking module for analyzing target compounds in complex matrices |
DESKTOP WORKSTATION | Desktop workstation Suitable to GC-MS - inclusive in the scope of supply | Yes |
Graphics Card Memory Size | 0, 1, 2, 4 | |
ONLINE UPS | UPS Provided in the Scope of Supply | Yes |
Rating of online UPS(KVA) | 5.0 KVA, 7.5 KVA, 10 KVA, NA | |
Backup time of online UPS | 60.0, 120.0 | |
GENERIC | Built-in Display | Yes, No |
Connectivity interface | USB, WIFI, ETHERNET, Bluetooth, RS 232, RJ-45 | |
Provision to split the sample from single injection to two different detectors and get results for two detectors simultaneously | Yes, No | |
Back Flash - inclusive in the scope of supply | Post Column Back Flash, Mid Column Back Flash, NA | |
ACESSORIES | Gas Supply inclusive in the scope of supply | Nitrogen, Hydrogen, Helium, Zero Air |
Number of Cyliners of each type supplied - inclusive in the scope of supply | 2 | |
Gas cylinders certified by PESO | Yes | |
Regulator - inclusive in the scope of supply | Dual Stage Regulator | |
"Other Accessories inclusive in the scope of supply (Select only those Accessories which are part of supply)" | Gas purification panel for all Gases with moisture trap, Hydrocarbon trap & Oxygen trap, Syringe for manual injection x 05 nos, Auto Sampler Vials x 1000 vials with screw cap, Graphite ferrules x 10 nos, Vessel ferrules x 10 nos, Septa for injector x 100 nos, Inlet liner for splitles x 4 nos, Inlet liner for split x 4 nos, O-rings x 10 nos. of each type, Column Cutter x 2 nos, Gas tube Cutter x 2 nos, Tool Kit x 1 no., User / Technical / Maintenance Manual, Standard Samples - NIST / IPL Traceable, Direct Injection Port, Splitter, Multishot Pyrolyser, Singleshot Pyrolyser, Syringe for Auto Sampler injection x 10 nos, EI filaments x 5 nos | |
ENVIRONMENTAL | Operating Humidity (RH) (%) at 40 degree C | 95.0 - 95.0 Or higher |
INSTALLATION & COMMISSIONING / WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on equipment | 1, 2, 3, 5 |
Number of days onsite training provided | 3, 5, 7, 10 | |
CERTIFICATION / TEST REPORT | "Availability of UL/CE Certificate as per EN 61010- 1:2010 Safety requirements for electrical equipment for measurement,control, and laboratory use General requirements" | Yes |
Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | Yes | |
Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | Yes, No | |