Catalogue values for ICU Ventilators (V2)
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
PRODUCT INFORMATION | Patient Type | Adult & pediatric |
Type of technology | Compressor | |
Tidal volume in ml | (20 ml or less) to (1500 ml or more) | |
Respiration rate, breaths/min | up to 80 or more, upto 150 or more | |
Trigger mechanism | Both pressure and flow | |
Inspiratory flow rate, L/min | up to 150 or more, upto 200 or more | |
Inspiratory pressure, cm H2O | upto 80 or more | |
IE ratio | 1:10 to 4:1 | |
Sigh breath function | Yes, No | |
Control panel lock | Yes | |
Facility for double lung ventilation | Yes | |
PATIENT ASSESSMENT TOOLS | Number of loops | 2 loops (P-V, F-V) with facility of saving of 1 Loops for reference |
Lung recruitment tools (PV loops) / OLT (Open lung tools) | Yes | |
Lung mechanics visualization tool | Yes | |
Capnography/CO2 monitoring | Yes | |
Esophageal / transpulmonary pressure monitoring | Yes | |
APRV(Airway Pressure Release Ventilation) /Bivent/Bi level | Yes | |
MMV(Mandatory Minute Ventilation) + PSV (Pressure Support Ventilation) /ASV( Adaptive Support Ventilation ) or ALPV(Adaptive Lung Protection Ventilation) or Automode | Yes | |
PRVC(Pressure-Regulated Volume Control ventilation)/Auto flow | Yes | |
High flow oxygen therapy | Yes | |
Automatic weaning system provided | Yes | |
Volume Support | Yes | |
RSBI (Rapid Shallow Breathing Index) | Yes | |
Expiratory Time constant ( Tcexp) | Yes | |
Compliance static and/or dynamic | Yes | |
Inspired and/or expired resistance | Yes | |
Occlusion pressure | Yes | |
Inspiratory and expiratory hold | Yes | |
Spontaneous frequency | Yes | |
Total peep, intrinsic peep, extrinsic peep | Yes | |
Expiratory block is autoclavable and no routine calibration required | Yes | |
Auto compensation for ET Tube | Yes | |
Facility for automatic compliance & Leakage compensation for circuit available | Yes | |
RS 232 out put port | Yes | |
USB/Memory card | Yes | |
Availability of Remote alarm/display port | Yes | |
Facility to Report (vent alarms and patient status) | Yes | |
HL/7 Compliant out put | Yes, No | |
Type of coloured display | LED/LCD/TFT | |
Display should be touch screen | Yes | |
NIV (Non Invasive Ventilation) to be possible in all modes of ventilation available | Yes, NIV available in pressure targeted modes only | |
Minimum Size of display (±2%) | 15.0 | |
Graphic display have automatic scaling facility for waves | Yes | |
HARDWARE AND ELECTRICAL REQUIREMENTS | Provision of UPS | Yes, No |
Backup time in hrs | 1, 2 | |
Built in air source | Air compressor (external) | |
Availability of stand alone compressor | Yes | |
Compressor shall provide an oil free Medical air, with Peak output flow should be minimum 150 LPM and Air quality complying with ISO compressed air purity class | Yes | |
Seller shall ensure compatibility of compressor with ventilator | Yes | |
Nebulizer with particle size less than 3 micron | Yes | |
ADDITIONAL ACCESSORIES | NV mask(Re usable)-Small,medium and large | 2 Each |
Humidifier: Servo controlled with digital monitoring of inspired gas temperature with heating wire | 1, 2, NA | |
WARRANTY | Warranty (years) | 5 Or higher |