Catalogue values for ICU Ventilators (V2)

Category Specification Bid Requirement (Allowed Values)
PRODUCT INFORMATION Patient Type Adult & pediatric
Type of technology Compressor
Tidal volume in ml (20 ml or less) to (1500 ml or more)
Respiration rate, breaths/min up to 80 or more, upto 150 or more
Trigger mechanism Both pressure and flow
Inspiratory flow rate, L/min up to 150 or more, upto 200 or more
Inspiratory pressure, cm H2O upto 80 or more
IE ratio 1:10 to 4:1
Sigh breath function Yes, No
Control panel lock Yes
Facility for double lung ventilation Yes
PATIENT ASSESSMENT TOOLS Number of loops 2 loops (P-V, F-V) with facility of saving of 1 Loops for reference
Lung recruitment tools (PV loops) / OLT (Open lung tools) Yes
Lung mechanics visualization tool Yes
Capnography/CO2 monitoring Yes
Esophageal / transpulmonary pressure monitoring Yes
APRV(Airway Pressure Release Ventilation) /Bivent/Bi level Yes
MMV(Mandatory Minute Ventilation) + PSV (Pressure Support Ventilation) /ASV( Adaptive Support Ventilation ) or ALPV(Adaptive Lung Protection Ventilation) or Automode Yes
PRVC(Pressure-Regulated Volume Control ventilation)/Auto flow Yes
High flow oxygen therapy Yes
Automatic weaning system provided Yes
Volume Support Yes
RSBI (Rapid Shallow Breathing Index) Yes
Expiratory Time constant ( Tcexp) Yes
Compliance static and/or dynamic Yes
Inspired and/or expired resistance Yes
Occlusion pressure Yes
Inspiratory and expiratory hold Yes
Spontaneous frequency Yes
Total peep, intrinsic peep, extrinsic peep Yes
Expiratory block is autoclavable and no routine calibration required Yes
Auto compensation for ET Tube Yes
Facility for automatic compliance & Leakage compensation for circuit available Yes
RS 232 out put port Yes
USB/Memory card Yes
Availability of Remote alarm/display port Yes
Facility to Report (vent alarms and patient status) Yes
HL/7 Compliant out put Yes, No
Type of coloured display LED/LCD/TFT
Display should be touch screen Yes
NIV (Non Invasive Ventilation) to be possible in all modes of ventilation available Yes, NIV available in pressure targeted modes only
Minimum Size of display (±2%) 15.0
Graphic display have automatic scaling facility for waves Yes
HARDWARE AND ELECTRICAL REQUIREMENTS Provision of UPS Yes, No
Backup time in hrs 1, 2
Built in air source Air compressor (external)
Availability of stand alone compressor Yes
Compressor shall provide an oil free Medical air, with Peak output flow should be minimum 150 LPM and Air quality complying with ISO compressed air purity class Yes
Seller shall ensure compatibility of compressor with ventilator Yes
Nebulizer with particle size less than 3 micron Yes
ADDITIONAL ACCESSORIES NV mask(Re usable)-Small,medium and large 2 Each
Humidifier: Servo controlled with digital monitoring of inspired gas temperature with heating wire 1, 2, NA
WARRANTY Warranty (years) 5 Or higher