Catalogue values for Fourier Transform Infra Red (FTIR) Spectrometer

Category Specification Bid Requirement (Allowed Values)
OPTICAL PERFORMANCE Wavenumber- Maximum (Per cm) 6000, 7800
Wavenumber - Minimum (Per cm) 350, 600
Spectral (Wavenumber) Resolution (Per cm) 0.07, 0.25, 0.5, 0.8, 0.85, 0.9, 1, 2
Spectral (Wave Number) Accuracy (Per cm) (+/-) 0.01, 0.05, 0.1
Beam Splitter KBr, ZnSe
Automatic beam splitter changeover for MIR to NIR or MIR to FIR without manual intervention Yes
Mirror Coating Aluminium, Gold
Peak to Peak Signal to Noise Ratio for 1 minute scan 20000:1 or more, 30000:1 or more, 40000:1 or more, 45000:1 or more, 50000:1 or more, 55000:1 or more
DETECTOR Detector Type Deuterated L-Alanine doped Triglycine Sulphate (DLaTGS), Deuterated Triglycine Sulphate (DTGS), Broad band MCT, W-MCT, M-MCT, N-MCT, InGAs, InSb, LiTaO3, PAS, Si bolometer
Detector having temperature control Yes
LIGHT SOURCE Light Source Mid IR Source, White Light Souce, Steady state infrared emitter, High intensity Ceramic light source, Halogen lamp, Water cooled mercury light source
SAMPLE TYPE / ANALYSIS FACILITY Analyze Liquid samples Yes
Analyze Solid samples Yes
Analyze Powder samples Yes
Analyze thin films on metals samples Yes, No
Analyze Gas samples No
SOFTWARE Customizable Report Template Yes
FTIR LIBRARIES FTIR Georgian Crime Library - inclusive in the scope of supply No, Yes
FTIR Canadian Forensic Library - inclusive in the scope of supply No
FTIR Sadler Drug Library - inclusive in the scope of supply No
FTIR Polymers & Polymer additives Library, nclusive in the scope of supply No, Yes
FTIR Surfactants Library, inclusive in the scope of supply No
FTIR Minerals & inorganic CompoundsLibrary, inclusive in the scope of supply No, Yes
FTIR Organic Compounds Library, inclusive in the scope of supply No, Yes
FTIR Drugs Library, inclusive in the scope of supply No, Yes
FTIR Adhesives & Sealants Library, inclusive in the scope of supply No
FTIR Cosmetics & Personal products Library, inclusive in the scope of supply No
ACCESSORIES Suitable Hydraulic Press 15 Ton for pellet/sample preparation - inclusive in the scope of supply No
Suitable Mortar and pestle - inclusive in the scope of supply No
Suitable Dies - inclusive in the scope of supply No
Suitable KBr set for Pellets - inclusive in the scope of supply No
KBr Powder - inclusive in the scope of supply No
Suitable Pellet holder - inclusive in the scope of supply No
Types of Suitable ATR - inclusive in the scope of supply Yes, No
ATR crystal - inclusive in the scope of supply Diamond
Suitable Universal liquid cell holder - inclusive in the scope of supply No
Suitable Dry Box - inclusive in the scope of supply No
Suitable Replacement dessicant (rechargeable) - inclusive in the scope of supply No, Yes
Suitable Liquid Cell Window - inclusive in the scope of supply No
Liquid Cell Window material NA
Shape of the Cell Window NA
Suitable Gas cell - inclusive in the scope of supply No
GENERIC FTIR and Sample compartment having provision for purging with gas No, Yes
FTIR provided with in-built air drier for humidity and vapour protection No
Connectivity USB, WIRELESS, WIFI, ETHERNET, Blue tooth, RS 232
OPERATING CONDITIONS Minimum Operating Temperature 0, 5, 10, 20
Maximum Operating Temperature 30, 40, 45, 50, 55, 60
Operating Humidity (RH) (%) at 40 degree C 65.0, 90.0
WARRANTY / AVAILABILITY OF SPARES / TRAINING Warranty on equipment 1.0, 3.0
Warranty on Source 10.0, 5.0
Warranty on interferometer 10.0
Warranty on Laser 10.0
Onsite training 5, 7, 10, 15
CERTIFICATION Certifications ISO, CE
Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements Yes, No
Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use Yes, No
Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification No
If Yes, Test Report to be furnished to the Buyer on demand Yes, NA