Catalogue values for Auto Refractometer (V2)
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
GENERAL | Product Description | Auto Refractometer |
PRODUCT INFORMATION | Eye Fixation | Auto-fogging system |
Alignment | Pupil rings | |
Change of magnification is auto until the image comes into focus | Yes | |
Spherical refractive measurable range (when VD=12 mm) | -20D to +20D | |
Stepup increments of sphere measurement | 0.125D | |
Cylindrical refractive measurable range | -10D to +10D | |
Stepup increments of cylinder measurement | 0.125D | |
Refractive measurable axis range (degree) | <=180 | |
Stepup increments of axis measurement | 1 degree | |
Pupil distance, maximum (PD) (mm) | 85.0 | |
Stepup increments of PD (mm) | +/-1 | |
Visual or Vertex Distance (VD) (mm) | 0, 12, 13.5 | |
Cylinder form | Yes | |
Minimum pupil dia (mm) | 2.0 | |
Measurement values for each eye | 10 | |
Standard value for each eye | 1 | |
Display Type | LCD | |
Display Size | 5 inch | |
Vertically adjustable chin rest, range (mm) | +/-25 | |
Printer type | Built in printer | |
Printing details | Measurements of each eye, date, time, Serial number and eye diagram | |
Built-in lamp for illumination, available | Yes | |
Power supply | 100~240V; 50Hz~60Hz | |
Power saver mode | Entire system viz. Monitor, lamp and motor go into sleep mode when operation is discontinued for 5 minutes (approx) | |
Motorized table | Yes | |
CERTIFICATIONS | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
Availability of valid Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes | |
Certification for manufacturing unit | ISO:13485 (Latest) | |
Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date | Yes | |
Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes | |
Electrical Safety | IEC 60601-1-2 or Equivalent BIS | |
WARRANTY | Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 5 Or higher |