Catalogue values for Ultrasound Machine (V2)
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
GENERAL | Product Description | Ultrasound Machine |
Types of Probes & Features | Applications | Abdomen, Abdominal vascular, Urology, OB/GYN, Neonatal brain, Small parts (breast, thyroid, scrotal, prostate, musculoskeletal) |
Convex Probe | 2.5 to 5 MHz (±2) | |
Microconvex probe | 4 to 8 MHz (±2), Not Provided | |
Linear array probe | 5 to 10 MHz (±2), 3 to 12 MHz (±2), 6 to 15 MHz (±2), 6 to 18 MHz (±2) | |
TVS Probe | 4 to 9 MHz (±2) | |
Cardiac Probe | 1 to 5 MHz (±2), 5 to 9 MHz (±2), Not Provided | |
Hockey Stick Probe | 6 to 17 MHz (±2), 6 to 13 MHz (±2), 8 to 20 MHz (±2), Not Provided | |
Volume Convex Probe | 2.5 to 5 MHz (±2) | |
Volume Linear Probe | 3 to 12 MHz (±2), 6 to 15 MHz (±2), 6 to 18 MHz (±2), 5 to 10 MHz (±2), Not Provided | |
Volume TVS Probe | 4 to 9 MHz (±2), Not Provided | |
Endorectal Probe | 5 to 9 MHz (±2), Not Provided | |
E-Breast Elastography technique | Yes, No | |
E-Thyroid Elastography technique | Yes, No | |
Liver Elastography technique | Yes, No | |
Gel Warmer | Yes, No | |
Dedicated 3-D/4-D | Yes | |
Scan Modes | B-mode | Yes |
M-mode | Yes | |
Contrast harmonic imaging with angio mode | Yes, No | |
Tissue harmonic imaging mode | Yes | |
Provision of Harmonic imaging in Linear Probes | Yes | |
Color Doppler imaging (CDI) | Yes | |
Color Doppler 3-D/4-D option | Yes | |
Power Doppler imaging (PDI) | Yes | |
Power Doppler 3-D/4-D option | Yes | |
Continuous wave | Yes | |
Pulsed wave | Yes | |
Duplex mode | Yes | |
Triplex mode | Yes, No | |
Tissue Doppler imaging | Yes, No | |
Contrast-enhanced ultrasound (CEUS) option | Yes, No | |
Tissue synchronization, tissue velocity | Yes, No | |
Gain control in two dimensions for additional level of flexibility for image quality control | Yes | |
Provision for higher resolution | Real time high frequency 2D | |
Probes supplied shall be in latest generation bandwidth | Yes | |
Provision for higher sensitivity for low frequency doppler in all probes | Yes, No | |
Number of channels in color doppler system | 50000 or more (if channels are being designated as digital channels), 300000 or more (if channels are being designated as digital channels), 1000000 or more (if channels are being designated as digital channels), 2500000 or more (if channels are being designated as digital channels), 4000000 or more (if channels are being designated as digital channels), 128 or more (if channels are being designated as physical channels) | |
Live 3 D imaging in color doppler system | Yes, No | |
Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval | Yes, No | |
Number of grey shades for sharp contrast resolutions in db | 256 or more | |
High-definition acoustic zoom provision for enlarging sections of 2D and Colour flow images with more acoustic information for greater clarity and detail while maintaining an optimal frame rate | Yes | |
Colour Flow Imaging Option | Yes | |
Provision of Tissue Colorization (B-Colour) for improved contrast resolution | Yes, No | |
All application packages should be inbuilt into the system | Yes | |
Cine Loop Features | Yes | |
Cine Loop Memory in MB | >=100 | |
Frame grabber facility for post analysis | Yes, No | |
User defined system and application presets for multi-user department | Yes | |
Number of transducer/probe ports | 4 or More | |
System should be capable of generating real time live 3- D images | Yes, No | |
Contrast agent imaging | Yes, No | |
Color Doppler system with all applications package, quad loop for serial studies with high frame rate, harmonic imaging capability in all modes and digital storage and retrieval | Yes | |
Image Display and Processing | DICOM 3 Compliant | Yes |
Adjustable transmit focus | Yes | |
Automated B-mode (2-D) image | Yes | |
Automated CDI(Color Doppler Imaging) image optimization | Yes, No | |
Automated PW Doppler image optimization | Yes, No | |
Dynamic receive focus | Yes | |
Grayscale levels | 256 | |
Image magnification (zoom) | Yes | |
Real-time image | Yes | |
Frozen image | Yes | |
Touch Screen | Yes | |
Monitor shall have high-resolution | Yes | |
Provision of Tilt or Swivel monitor should be able to view in all angles and all light conditions | Yes | |
Type of Display | LCD/LED | |
Monitor display size (inch) | 21 or more Or higher | |
Split screen | Twin View, Twin View and Live Quad Screen, No | |
Single/dual monitors | Single, Single with optional second console, Dual (image display and touch panel) | |
Depth, cm | ≥30 | |
Pre processing | Yes | |
Post processing | Yes | |
Spectrum analyzer | Yes | |
Digital Image Storage | Yes | |
Cine loop playback | Yes | |
Max number frames | >=1000 | |
Digital storage hard drive, GB | >=500 | |
Speckle-tracking strain and strain rate | Yes | |
Exam protocols | Yes | |
Digital calipers | Yes | |
Exam presets | Yes | |
Customizable presets | Yes | |
On-screen annotation | Yes | |
DVD/CD Writer/ USB | Yes | |
Flash Memory | Yes, No | |
USB connector | Yes | |
Facility for high definition digital acquisition, review and editing of complete patient studies | Yes | |
Type of processor (in PC) | i7 or latest | |
RAM capacity of PC in GB (hint: In case PC not provided put zero) | ≥ 4 | |
Processor speed in GHz (PC) | ≥ 3 | |
Frame Grabber incorporated(PC) | Yes, No | |
Connectivity between USG and PC | Yes | |
PC with all software inclusive interfaced with USG machine | Yes | |
Laser Color printer provided | Yes, No | |
Inkjet/Thermal printer provided | Yes | |
CD/DVD/USB Produced should be playable on any system | Yes | |
Glossy Colour Print Paper supplied | 500 sheets, 1000 sheets, NA | |
Dicom Interfacability Dicom to PACS | Yes | |
Interfacability Dicom to RIS | Yes, No | |
USG sytem supplied should be upgradable to next generation system on site | Yes | |
Power Requirements | Power input, VAC | 220-240VAC, 50Hz fitted with Indian plug |
Resettable over current breaker shall be fitted for protection | Yes, No | |
Power backup | UPS | |
Back up time in minutes (minimum) | 30, 60, 90, 120 | |
Suitable Servo Controlled Stabilizer / CVT should be provided | Yes, No | |
Certifications | Compliance to Medical Device Rules (MDR) 2017 as amended till date | Yes |
Availability of valid drug/Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date | Yes | |
Certification for manufacturing unit | ISO:13485 (Latest) | |
Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date | Yes | |
Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement | Yes | |
Electrical Safety standards | IEC-60601-1-1 or equivalent BIS | |
Warranty and Maintenance | Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 3, 5 Or higher |
Miscellaneous Requirements | OEM/Reseller shall ensure uninterrupted availability of all spares, consumables and reagents for atleast 10 years | Yes |
Availability of toll free facility for technical support maintened by OEM or authorized agencies | Yes | |
User/Technical/Maintenance manuals to be provided in English in hard and soft copy | Yes | |
Details of equipments and procedures required for local calibration and routine maintenance to be provided and advanced maintenance task documentation also to be furnished | Yes | |
Supplier to perform installation, safety and operation checks before handover | Yes | |
Training of users in operation and basic maintenance shall be provided | Yes | |
Contact details of manufacturer, supplier and local service agent to be provided | Yes | |
Certificate of calibration from the manufacturer | Yes | |