Catalogue values for Fourier Transform Infra Red (FTIR) Spectrometer

Category Specification Bid Requirement (Allowed Values)
OPTICAL PERFORMANCE Wavenumber- Maximum (Per cm) 6000
Wavenumber - Minimum (Per cm) 600
Spectral (Wavenumber) Resolution (Per cm) 0.5
Spectral (Wave Number) Accuracy (Per cm) (+/-) 0.05
Beam Splitter ZnSe
Automatic beam splitter changeover for MIR to NIR or MIR to FIR without manual intervention No
Mirror Coating Gold
Peak to Peak Signal to Noise Ratio for 1 minute scan 55000:1 or more
DETECTOR Detector Type Deuterated L-Alanine doped Triglycine Sulphate (DLaTGS)
Detector having temperature control Yes
LIGHT SOURCE Light Source High intensity Ceramic light source
SAMPLE TYPE / ANALYSIS FACILITY Analyze Liquid samples Yes
Analyze Solid samples Yes
Analyze Powder samples Yes
Analyze thin films on metals samples Yes
Analyze Gas samples Yes
SOFTWARE Customizable Report Template Yes
FTIR LIBRARIES FTIR Georgian Crime Library - inclusive in the scope of supply No
FTIR Canadian Forensic Library - inclusive in the scope of supply No
FTIR Sadler Drug Library - inclusive in the scope of supply No
FTIR Polymers & Polymer additives Library, nclusive in the scope of supply No
FTIR Surfactants Library, inclusive in the scope of supply No
FTIR Minerals & inorganic CompoundsLibrary, inclusive in the scope of supply No
FTIR Organic Compounds Library, inclusive in the scope of supply No
FTIR Drugs Library, inclusive in the scope of supply No
FTIR Adhesives & Sealants Library, inclusive in the scope of supply No
FTIR Cosmetics & Personal products Library, inclusive in the scope of supply No
ACCESSORIES Suitable Hydraulic Press 15 Ton for pellet/sample preparation - inclusive in the scope of supply No
Suitable Mortar and pestle - inclusive in the scope of supply No
Suitable Dies - inclusive in the scope of supply No
Suitable KBr set for Pellets - inclusive in the scope of supply No
KBr Powder - inclusive in the scope of supply No
Suitable Pellet holder - inclusive in the scope of supply Yes
Types of Suitable ATR - inclusive in the scope of supply No
ATR crystal - inclusive in the scope of supply NA
Suitable Universal liquid cell holder - inclusive in the scope of supply Yes
Suitable Dry Box - inclusive in the scope of supply Yes
Suitable Replacement dessicant (rechargeable) - inclusive in the scope of supply Yes
Suitable Liquid Cell Window - inclusive in the scope of supply Yes
Liquid Cell Window material CaF2
Shape of the Cell Window Rectangular
Suitable Gas cell - inclusive in the scope of supply No
GENERIC FTIR and Sample compartment having provision for purging with gas No
FTIR provided with in-built air drier for humidity and vapour protection No
Connectivity ETHERNET
OPERATING CONDITIONS Minimum Operating Temperature 20.0
Maximum Operating Temperature 60.0
Operating Humidity (RH) (%) at 40 degree C 95.0
WARRANTY / AVAILABILITY OF SPARES / TRAINING Warranty on equipment 1.0
Warranty on Source 5.0
Warranty on interferometer 5.0
Warranty on Laser 5.0
Onsite training 5.0
CERTIFICATION Certifications ISO, CE
Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements No
Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use No
Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification No
If Yes, Test Report to be furnished to the Buyer on demand NA