Catalogue values for Fourier Transform Infra Red (FTIR) Spectrometer
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
OPTICAL PERFORMANCE | Wavenumber- Maximum (Per cm) | 6000, 7000, 7100, 7500, 7800, 8000, 12000 |
Wavenumber - Minimum (Per cm) | 350, 375, 450, 600 | |
Spectral (Wavenumber) Resolution (Per cm) | 0.07, 0.25, 0.5, 0.8, 0.85, 0.9, 1, 2 | |
Spectral (Wave Number) Accuracy (Per cm) (+/-) | 0.01, 0.05, 0.1 | |
Beam Splitter | KBr, ZnSe | |
Automatic beam splitter changeover for MIR to NIR or MIR to FIR without manual intervention | Yes, No | |
Mirror Coating | Aluminium, Gold | |
Peak to Peak Signal to Noise Ratio for 1 minute scan | 20000:1 or more, 30000:1 or more, 40000:1 or more, 45000:1 or more, 50000:1 or more, 55000:1 or more | |
DETECTOR | Detector Type | Deuterated L-Alanine doped Triglycine Sulphate (DLaTGS), Deuterated Triglycine Sulphate (DTGS), Broad band MCT, W-MCT, M-MCT, N-MCT, InGAs, InSb, LiTaO3, PAS, Si bolometer |
Detector having temperature control | Yes, No | |
LIGHT SOURCE | Light Source | Mid IR Source, White Light Souce, Steady state infrared emitter, High intensity Ceramic light source, Halogen lamp, Water cooled mercury light source |
SAMPLE TYPE / ANALYSIS FACILITY | Analyze Liquid samples | Yes, No |
Analyze Solid samples | Yes, No | |
Analyze Powder samples | Yes, No | |
Analyze thin films on metals samples | Yes, No | |
Analyze Gas samples | Yes, No | |
SOFTWARE | Customizable Report Template | Yes, No |
FTIR LIBRARIES | FTIR Georgian Crime Library - inclusive in the scope of supply | Yes, No |
FTIR Canadian Forensic Library - inclusive in the scope of supply | Yes, No | |
FTIR Sadler Drug Library - inclusive in the scope of supply | Yes, No | |
FTIR Polymers & Polymer additives Library, nclusive in the scope of supply | Yes, No | |
FTIR Surfactants Library, inclusive in the scope of supply | Yes, No | |
FTIR Minerals & inorganic CompoundsLibrary, inclusive in the scope of supply | Yes, No | |
FTIR Organic Compounds Library, inclusive in the scope of supply | Yes, No | |
FTIR Drugs Library, inclusive in the scope of supply | Yes, No | |
FTIR Adhesives & Sealants Library, inclusive in the scope of supply | Yes, No | |
FTIR Cosmetics & Personal products Library, inclusive in the scope of supply | Yes, No | |
ACCESSORIES | Suitable Hydraulic Press 15 Ton for pellet/sample preparation - inclusive in the scope of supply | Yes, No |
Suitable Mortar and pestle - inclusive in the scope of supply | Yes, No | |
Suitable Dies - inclusive in the scope of supply | Yes, No | |
Suitable KBr set for Pellets - inclusive in the scope of supply | Yes, No | |
KBr Powder - inclusive in the scope of supply | Yes, No | |
Suitable Pellet holder - inclusive in the scope of supply | Yes, No | |
Types of Suitable ATR - inclusive in the scope of supply | Yes, No | |
ATR crystal - inclusive in the scope of supply | Diamond, ZnSe, Germanium (Ge), NA | |
Suitable Universal liquid cell holder - inclusive in the scope of supply | Yes, No | |
Suitable Dry Box - inclusive in the scope of supply | Yes, No | |
Suitable Replacement dessicant (rechargeable) - inclusive in the scope of supply | Yes, No | |
Suitable Liquid Cell Window - inclusive in the scope of supply | Yes, No | |
Liquid Cell Window material | KBr, ZnSe, CaF2, CsI, NaCl, KrS5, NA | |
Shape of the Cell Window | Round, Rectangular, NA | |
Suitable Gas cell - inclusive in the scope of supply | Yes, No | |
GENERIC | FTIR and Sample compartment having provision for purging with gas | Yes, No |
FTIR provided with in-built air drier for humidity and vapour protection | Yes, No | |
Connectivity | USB, WIRELESS, WIFI, ETHERNET, Blue tooth, RS 232 | |
OPERATING CONDITIONS | Minimum Operating Temperature | 0, 5, 10, 20 |
Maximum Operating Temperature | 30, 40, 45, 50, 55, 60 | |
Operating Humidity (RH) (%) at 40 degree C | 65, 70, 75, 80, 85, 90, 95 | |
WARRANTY / AVAILABILITY OF SPARES / TRAINING | Warranty on equipment | 1, 2, 3, 5 |
Warranty on Source | 1, 2, 3, 5, 10 | |
Warranty on interferometer | 1, 2, 3, 5, 10 | |
Warranty on Laser | 1, 2, 3, 5, 10 | |
Onsite training | 5, 7, 10, 15 | |
CERTIFICATION | Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements | Yes, No |
Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use | Yes, No | |
Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | Yes, No | |