Catalogue values for High quality water purification system
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
Performance Parameters | Purpose | Lab applications for HPLC/Ion Chromatography / LCMS/Titration |
Prefilteration RO Unit supplied | Yes Or higher | |
In put quality in TDS for prefileteration unit | To be less than 2000 TDS | |
Out put quality from prefilteration RO Unit in TDS | 300< | |
Out put flow rate for pre filteration unit | Minimum 50 litre per hour to be provided | |
In case pre filteration unit provided capacity of water storage tank from pre filteration unutb | Minimum 50 litre | |
Type of water produced by system | Type1 and Type 2 | |
Resistivity of water out put | 10-15 mega ohms .cms or better @25 degree celsius (For Type I), 18 mega ohms.cms or better @25 degree celcius 9 For Type 2) | |
Toc (ppb) | <05 (Type 1) | |
Display provided | LCD | |
Active carbon beads Electro Deionization Module (EDI) | Yes Or higher | |
Provision of UV in storage tank | Yes Or higher | |
Miscellaneous Parameters | For easy understanding of main operations Built in quick reference guide in system | Yes Or higher |
Number of installations of the Machine machines in Central /State/PSU Govt Hospitals( Note: Seller should supply a performace certificate of the device to the buyer if demanded after placement of order) | No installations, Upto 3, More than 3 | |