Catalogue values for High quality water purification system

Category Specification Bid Requirement (Allowed Values)
Performance Parameters Purpose Lab applications for HPLC/Ion Chromatography / LCMS/Titration
Prefilteration RO Unit supplied Yes Or higher
In put quality in TDS for prefileteration unit To be less than 2000 TDS
Out put quality from prefilteration RO Unit in TDS 300<
Out put flow rate for pre filteration unit Minimum 50 litre per hour to be provided
In case pre filteration unit provided capacity of water storage tank from pre filteration unutb Minimum 50 litre
Type of water produced by system Type1 and Type 2
Resistivity of water out put 10-15 mega ohms .cms or better @25 degree celsius (For Type I), 18 mega ohms.cms or better @25 degree celcius 9 For Type 2)
Toc (ppb) <05 (Type 1)
Display provided LCD
Active carbon beads Electro Deionization Module (EDI) Yes Or higher
Provision of UV in storage tank Yes Or higher
Miscellaneous Parameters For easy understanding of main operations Built in quick reference guide in system Yes Or higher
Number of installations of the Machine machines in Central /State/PSU Govt Hospitals( Note: Seller should supply a performace certificate of the device to the buyer if demanded after placement of order) No installations, Upto 3, More than 3