Catalogue values for Ultrasound Machine (V2)

Category Specification Bid Requirement (Allowed Values)
GENERAL Product Description Ultrasound Machine
Types of Probes & Features Applications Abdomen, Abdominal vascular, Urology, OB/GYN, Small parts (breast, thyroid, scrotal, prostate, musculoskeletal)
Convex Probe 2.5 to 5 MHz (±2)
Microconvex probe Not Provided
Linear array probe 3 to 12 MHz (±2)
TVS Probe 4 to 9 MHz (±2)
Cardiac Probe Not Provided
Hockey Stick Probe Not Provided
Volume Convex Probe Not Provided
Volume Linear Probe Not Provided
Volume TVS Probe Not Provided
Endorectal Probe Not Provided
E-Breast Elastography technique No
E-Thyroid Elastography technique No
Liver Elastography technique No
Gel Warmer No
Dedicated 3-D/4-D Yes, No
Scan Modes B-mode Yes
M-mode Yes
Contrast harmonic imaging with angio mode Yes
Tissue harmonic imaging mode Yes
Provision of Harmonic imaging in Linear Probes Yes
Color Doppler imaging (CDI) Yes
Color Doppler 3-D/4-D option Yes, No
Power Doppler imaging (PDI) Yes
Power Doppler 3-D/4-D option Yes, No
Continuous wave No
Pulsed wave Yes
Duplex mode Yes
Triplex mode Yes, No
Tissue Doppler imaging Yes, No
Contrast-enhanced ultrasound (CEUS) option Yes, No
Tissue synchronization, tissue velocity Yes, No
Gain control in two dimensions for additional level of flexibility for image quality control Yes
Provision for higher resolution Real time high frequency 2D
Probes supplied shall be in latest generation bandwidth Yes
Provision for higher sensitivity for low frequency doppler in all probes Yes
Number of channels in color doppler system 300000 or more (if channels are being designated as digital channels)
Live 3 D imaging in color doppler system Yes, No
Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval Yes, No
Number of grey shades for sharp contrast resolutions in db 256 or more
High-definition acoustic zoom provision for enlarging sections of 2D and Colour flow images with more acoustic information for greater clarity and detail while maintaining an optimal frame rate Yes
Colour Flow Imaging Option No, Yes
Provision of Tissue Colorization (B-Colour) for improved contrast resolution Yes
All application packages should be inbuilt into the system Yes
Cine Loop Features Yes
Cine Loop Memory in MB >=100
Frame grabber facility for post analysis Yes, No
User defined system and application presets for multi-user department Yes
Number of transducer/probe ports 4 or More
System should be capable of generating real time live 3- D images Yes, No
Contrast agent imaging Yes, No
Color Doppler system with all applications package, quad loop for serial studies with high frame rate, harmonic imaging capability in all modes and digital storage and retrieval No, Yes
Image Display and Processing DICOM 3 Compliant Yes, No
Adjustable transmit focus Yes
Automated B-mode (2-D) image Yes
Automated CDI(Color Doppler Imaging) image optimization Yes
Automated PW Doppler image optimization Yes
Dynamic receive focus Yes
Grayscale levels 256
Image magnification (zoom) Yes
Real-time image Yes
Frozen image Yes
Touch Screen Yes, No
Monitor shall have high-resolution Yes
Provision of Tilt or Swivel monitor should be able to view in all angles and all light conditions Yes
Type of Display LCD/LED
Monitor display size (inch) 21 or more Or higher
Split screen Twin View
Single/dual monitors Single
Depth, cm ≥30
Pre processing Yes
Post processing Yes
Selectable dynamic range (dB) >=200
Spectrum analyzer Yes
Digital Image Storage Yes
Cine loop playback Yes
Max number frames >=1000
Digital storage hard drive, GB >=500
Speckle-tracking strain and strain rate No
Exam protocols Yes
Digital calipers Yes
Exam presets Yes
Customizable presets Yes
On-screen annotation Yes
DVD/CD Writer/ USB Yes
Flash Memory No
USB connector Yes
Facility for high definition digital acquisition, review and editing of complete patient studies Yes
Type of processor (in PC) i7 or latest, NA
RAM capacity of PC in GB (hint: In case PC not provided put zero) ≥ 4, NA
Processor speed in GHz (PC) ≥ 3, NA
Frame Grabber incorporated(PC) Yes, No
Connectivity between USG and PC Yes, No
PC with all software inclusive interfaced with USG machine Yes, No
Laser Color printer provided Yes, No
Inkjet/Thermal printer provided Yes
CD/DVD/USB Produced should be playable on any system Yes
Glossy Colour Print Paper supplied 500 sheets, 1000 sheets, NA
Dicom Interfacability Dicom to PACS Yes, No
Interfacability Dicom to RIS Yes, No
USG sytem supplied should be upgradable to next generation system on site Yes
Power Requirements Power input, VAC 220-240VAC, 50Hz fitted with Indian plug
Resettable over current breaker shall be fitted for protection Yes
Power backup UPS
Back up time in minutes (minimum) 30.0
Suitable Servo Controlled Stabilizer / CVT should be provided Yes, No
Certifications Compliance to Medical Device Rules (MDR) 2017 as amended till date Yes
Availability of valid drug/Medical Device license for the product issued from the competent authority defined under Drugs and Cosmetic Act 1940 and Rules made there under as amended till date Yes
Certification for manufacturing unit ISO:13485 (Latest)
Availability of Test Report for each supplied batch/product as per Medical Device Rule (MDR) 2017 as amended till date Yes
Submission of all necessary certifications, licenses and test reports to the buyer at the time of bid submission or along with supplies as per buyer requirement Yes
Electrical Safety standards IEC-60601-1-1 or equivalent BIS
Warranty and Maintenance Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 5 Or higher
Miscellaneous Requirements OEM/Reseller shall ensure uninterrupted availability of all spares, consumables and reagents for atleast 10 years Yes
Availability of toll free facility for technical support maintened by OEM or authorized agencies Yes
User/Technical/Maintenance manuals to be provided in English in hard and soft copy Yes
Details of equipments and procedures required for local calibration and routine maintenance to be provided and advanced maintenance task documentation also to be furnished Yes
Supplier to perform installation, safety and operation checks before handover Yes
Training of users in operation and basic maintenance shall be provided Yes
Contact details of manufacturer, supplier and local service agent to be provided Yes
Certificate of calibration from the manufacturer Yes