Catalogue values for Frequency Counter as per IS 9000
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
FEATURES | Frequency Range | 2 mHz to 6 GHz |
Number of Channels | 2 | |
Type of Input Signal | Burst, AM, FM, Noise Signal | |
Resolution (Digit) | 12 | |
Input Impedance | 1 M Ohm and 50 Ohm (Selectable) | |
Trigger Slope | Positive, Negative | |
Aging Rate Per Year (As declared by the OEM) | 0.2 PPM or Better | |
Measurement | Frequency / Period, Time, Phase, Volt, Duty Cycle, Pulse, Rise Time | |
Number of Digits | Programmable | |
Time Base Type | OCXO | |
Gate Time | 100 micro sec to 1000 sec or better | |
Single Shot Time Resolution | 100.0 | |
Temperature Stability at 26 Degree C | .02 PPM | |
OPERATING CONDITION | Minimum Operating Temperature | 0.0 |
Maximum Operating Temperature | 55.0 | |
Operating Humidity (RH) @ 40 deg C | 80.0 | |
Storage Temperature (Degree C) | -30 to 50 degree C | |
POWER SUPPLY | Power Rating | 70, 80, 90 Or lower |
GENERIC | Indication | Frequency/Level/ Status Indication(Histogram) |
SERVICE | Warranty | 3 Or higher |
CERTIFICATION AND TEST REPORT DETAILS | Safety Certification (Certificate to be furnished when demanded by buyer) | UL, CSA |
conformity to Dry Heat for 16 hrs at a specified maximum temp in accordance with IS:9000/part-3/section-5/1977 latest | Yes | |
conformity to Cold Test for 4 hrs at a specified minimum temp in accordance with IS:9000/part-2/section-4/1977 latest | Yes | |
Conformity to Damp Heat (Cyclic)Test : for 2 Cycles of 24 hrs each at a temp of 40 degree C & 95% RH | Yes | |
Availability of Test Reports from Central Govt/NABL/ILAC approved accredited lab to prove conformity to the specification | Yes | |