Catalogue values for Breath Alcohol Analyzers
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
Battery of Analyzer | Chemistry of Rechargeable Battery | Li-ion |
Number of Spare Rechargeable Batteries supplied with the instrument | 0 | |
Power Source | Rechargeable Battery | |
Display / Indicator / Storage / Interface | Display Type | LCD |
Record Memory (Minimum) | 20000 | |
If yes, Capable to display the geolocation of test inspection / location | NA | |
GPS | No | |
Data Interface | USB | |
Camera | No | |
Record Entry | Through Alphanumeric Keyboard | |
If yes, Capable to capture the clear image of test target | NA | |
Environmental / Certification | Conformity to Dry Heat test at 50 degree C for 16 hours as per IS :9000 latest | Yes |
Conformity to Damp heat at 40 degree C - 95% RH (cyclic test): two cycles of 16+8 hours as per IS: 9000 latest | Yes | |
Conformity to Cold test at - 5 / -10 / - 15 / -20 degree C (as applicable) for 4 hours as per IS: 9000 latest | Yes | |
Availability of Test Reports from Central Govt / NABL approved / ILAC accredited Lab to prove conformity to the specification | Yes | |
Operating Conditions | Maximum Operating Temperature | 60.0 |
Minimum Operating Temperature | -10.0 | |
Relative Humidity Non Condensing (%) | 85 | |
Printer | Printer consists of Inbuilt Rechargeable Battery, Paper for offence printouts, Charger | Yes |
Printer | With | |
Type of Printer | Internal Thermal Printer | |
Statutory Requirements | If Yes, the equipment meets the requirements of the applicable latest Central Motor Vehicle Rules and all Govt notifications | Yes |
The equipment can be used as an evidential instrument for Traffic Law Enforcement, The instrument is hand held, portable and easy to operate | Yes | |
Technology / Performance | RoHS Compliant | Yes |
Sensor Technology | Electrochemical - Fuel Cell (alcohol specific) | |
Time required for re-calibration | 12.0 | |
Drift of the breath analyzer in 6 months period (mg / 100 ml BAC) | 1.0 - 1.05 Or lower | |
Measuring Range [mg / 100 ml BAC (Blood Alcohol Concentration)] | 0 to 600 | |
The Equipment meet RFI requirement as per EN:5008-1/1992(emission) and EN 50082-1:1997(immunity) or equivalent requirements | Yes | |
Warranty | Warranty | 1, 2, 3, 4, 5 Or higher |