Catalogue values for Non Contact Tonometer

Category Specification Bid Requirement (Allowed Values)
Performance Parameter Purpose It is used to measure the intraocular pressure (IOP) inside a patient's eyes by using a small puff of air which air flattens the eye's cornea in a non-invasive way.
The measurement of intraocular pressure (IOP) should be without actual eye contact Yes
Availability of APC (Automatic Puff Control) Yes
APC range setting APC 40/60
Type of display Type of display LED
Facility of touch display screen Yes
Type of measurement mode Both Single Button measurement mode & Touch triple measurement mode
Mode of connectivity USB
Mode of alignment & measurement Both automatic & Manual
Power Supply sources Single Phase (230 V, 50 Hz)
Facility of Auto-Shut Off power saving mechanism Yes
Facility of adjustable safety stopper Yes Or higher
Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 2, 3, 4, 5, 1
Comprehensive Maintenance Contract (CMC) including preventive maintenance, testing & calibration as per technical/ service /operational manual after satisfactory completion of Warranty period 1, 2, 3, 4, No CMC after warranty
Dimensions Parameter Pressure measurement range in mm Hg 1 to 60
Size of display in inch 7
Reports & Cerifications Certifications for product European CE (with approved document )