Catalogue values for Automated HPLC System for Separation of complex mixtures
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
Standard parameter | Instrument calibration standards | IEC 17025 |
Quaternary Gradient Solvent Delivery Unit with Degassing Unit | Components of unit | Quaternary Low Pressure Gradient pump & Serial Double Plunger |
Number of solvents delivery | 4 | |
Flow rate setting range in ml/min | 0.001 to 10 | |
Flow rate accuracy | ± 1 %, ± 0.1% | |
UV-Vis Detector | UV-Vis Detector provided | Yes Or higher |
Light source | Deuterium lamp | |
Spectrum Band width in nm | 4, 5, 6, 8, NA if not provided | |
Wavelength range in nm | 190 - 700, NA if not provided | |
Sampling Rate in Hz | 80, 90, 100, NA if not provided | |
The flow cell must be temperature Controlled | Yes, NA if not provided | |
Wavelength accuracy in nm | ± 1 | |
Wavelength reproducibility in nm | ± 0.1, NA if not provided | |
Drift in AU/Hour | less than 1 x 10-4 | |
Noise level in AU | ± 2.5 x 10-6 | |
Ability to monitor and quantitative 2 wavelengths simultaneously | Yes, NA if not provided | |
Linearity in AU | 2.5 or more, NA if not provided | |
Temperature control provided for Flow cell | Yes, NA if not provided | |
Temperature control range for flow cell in Degree Celsius | 5 degree above room temp to 50 degree, NA if not provided | |
flow cell path length in mm | 10, NA if not provided | |
Flow cell volume in ?l | 10 - 15, NA if not provided, 16 - 20 | |
Auto Sample Injector with cooler | Injection Method | Total volume sample injection |
Carry over of sample | 0.005% or lower | |
Injection precision in RSD | less than 0.20 % RSD, less than 0.50 % RSD | |
Software | Operating Windows | Win 10 or latest |
Photo Diode Array (PDA) Detector | Provision of Photo diode array | Yes Or higher |
PDA detector Wavelength range in nm | 190- 800 or better | |
Number of Diodes | 512, 1024, NA if not provided | |
PDA detector light source | Deuterium lamp and Tungsten lamp | |
PDA detector Wavelength accuracy in nm | ± 1 | |
Data Rate in Hz | 100, <100 | |
Noise in AU | less than ± 1 x 10-5 or lower | |
PDA detector Drift in AU/Hour | less than 1 x 10-3 | |
Temperature controlled flow cell in C | 5 degree above room temp to 50 degree, NA if not provided | |
Flow cell | Flow cell | |
Warranty & Reports | Warranty (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 2, 3, 4, 5, 1 Or higher |
Number of installations in Central /State/PSU Govt Hospitals | 0, <3, >3 | |
Availability of toll free facility for technical support maintained by OEM or authorized agencies | Yes Or higher | |