Catalogue values for Ultrasound Machine (V2)
Category | Specification | Bid Requirement (Allowed Values) |
---|---|---|
Types of Probes & Features | Applications | Abdomen, Abdominal vascular, Urology, OB/GYN, Neonatal brain, Small parts (breast, thyroid, scrotal, prostate, musculoskeletal) |
Convex Probe | 2.5 to 5 MHz (±2) | |
Microconvex probe | Not Provided | |
Linear array probe | 3 to 12 MHz (±2) | |
TVS Probe | 4 to 9 MHz (±2) | |
Cardiac Probe | Not Provided | |
Hockey Stick Probe | Not Provided | |
Volume Convex Probe | 2.5 to 5 MHz (±2), Not Provided | |
Volume Linear Probe | Not Provided | |
Volume TVS Probe | 4 to 9 MHz (±2) | |
Endorectal Probe | Not Provided | |
E-Breast Elastography technique | No | |
E-Thyroid Elastography technique | No | |
Liver Elastography technique | No | |
Gel Warmer | No | |
Dedicated 3-D/4-D | Yes | |
Scan Modes | B-mode | Yes |
Contrast harmonic imaging with angio mode | No | |
Provision of Harmonic imaging in Linear Probes | Yes | |
Color Doppler imaging (CDI) | Yes | |
Color Doppler 3-D/4-D option | Yes | |
Power Doppler imaging (PDI) | Yes | |
Power Doppler 3-D/4-D option | Yes | |
Continuous wave | Yes | |
Triplex mode | Yes | |
Tissue Doppler imaging | Yes | |
Contrast-enhanced ultrasound (CEUS) option | Yes | |
Tissue synchronization, tissue velocity | No | |
Provision for higher sensitivity for low frequency doppler in all probes | No | |
Number of channels in color doppler system | 4000000 or more (if channels are being designated as digital channels) | |
Live 3 D imaging in color doppler system | Yes | |
Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval | Yes | |
Provision of Tissue Colorization (B-Colour) for improved contrast resolution | Yes | |
Cine Loop Features | Yes | |
Frame grabber facility for post analysis | Yes | |
Number of transducer/probe ports | 4 or More | |
System should be capable of generating real time live 3- D images | Yes | |
Contrast agent imaging | Yes | |
Image Display and Processing | DICOM 3 Compliant | Yes |
Automated B-mode (2-D) image | Yes | |
Automated CDI(Color Doppler Imaging) image optimization | Yes | |
Automated PW Doppler image optimization | Yes | |
Touch Screen | Yes | |
Monitor display size (inch) | 21 or more Or higher | |
Split screen | Twin View and Live Quad Screen | |
Single/dual monitors | Dual (image display and touch panel) | |
Max number frames | >=1000 | |
Digital storage hard drive, GB | >=500 | |
DVD/CD Writer/ USB | Yes | |
Type of processor (in PC) | i7 or latest | |
RAM capacity of PC in GB (hint: In case PC not provided put zero) | ≥ 4 | |
Processor speed in GHz (PC) | ≥ 3 | |
Frame Grabber incorporated(PC) | No | |
Connectivity between USG and PC | No | |
PC with all software inclusive interfaced with USG machine | No | |
Laser Color printer provided | No | |
Inkjet/Thermal printer provided | No | |
CD/DVD/USB Produced should be playable on any system | No | |
Glossy Colour Print Paper supplied | NA | |
Dicom Interfacability Dicom to PACS | Yes | |
Interfacability Dicom to RIS | Yes | |
Power Requirements | Resettable over current breaker shall be fitted for protection | Yes |
Power backup | UPS | |
Back up time in minutes (minimum) | 30.0 | |
Suitable Servo Controlled Stabilizer / CVT should be provided | No | |
Warranty and Maintenance | Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) | 5 Or higher |