Catalogue values for Ultrasound Machine (V2)

Category Specification Bid Requirement (Allowed Values)
Types of Probes & Features Applications Abdomen, Abdominal vascular, Urology, OB/GYN, Neonatal brain, Small parts (breast, thyroid, scrotal, prostate, musculoskeletal)
Convex Probe 2.5 to 5 MHz (±2)
Microconvex probe Not Provided
Linear array probe 3 to 12 MHz (±2)
TVS Probe 4 to 9 MHz (±2)
Cardiac Probe Not Provided
Hockey Stick Probe Not Provided
Volume Convex Probe 2.5 to 5 MHz (±2), Not Provided
Volume Linear Probe Not Provided
Volume TVS Probe 4 to 9 MHz (±2)
Endorectal Probe Not Provided
E-Breast Elastography technique No
E-Thyroid Elastography technique No
Liver Elastography technique No
Gel Warmer No
Dedicated 3-D/4-D Yes
Scan Modes B-mode Yes
Contrast harmonic imaging with angio mode No
Provision of Harmonic imaging in Linear Probes Yes
Color Doppler imaging (CDI) Yes
Color Doppler 3-D/4-D option Yes
Power Doppler imaging (PDI) Yes
Power Doppler 3-D/4-D option Yes
Continuous wave Yes
Triplex mode Yes
Tissue Doppler imaging Yes
Contrast-enhanced ultrasound (CEUS) option Yes
Tissue synchronization, tissue velocity No
Provision for higher sensitivity for low frequency doppler in all probes No
Number of channels in color doppler system 4000000 or more (if channels are being designated as digital channels)
Live 3 D imaging in color doppler system Yes
Colour Doppler system with all application packages ,Quad loop for serial studies with high frame rate revies, harmonic imaging capability in all modes and integrated stress echo package digital storage and retrieval Yes
Provision of Tissue Colorization (B-Colour) for improved contrast resolution Yes
Cine Loop Features Yes
Frame grabber facility for post analysis Yes
Number of transducer/probe ports 4 or More
System should be capable of generating real time live 3- D images Yes
Contrast agent imaging Yes
Image Display and Processing DICOM 3 Compliant Yes
Automated B-mode (2-D) image Yes
Automated CDI(Color Doppler Imaging) image optimization Yes
Automated PW Doppler image optimization Yes
Touch Screen Yes
Monitor display size (inch) 21 or more Or higher
Split screen Twin View and Live Quad Screen
Single/dual monitors Dual (image display and touch panel)
Max number frames >=1000
Digital storage hard drive, GB >=500
DVD/CD Writer/ USB Yes
Type of processor (in PC) i7 or latest
RAM capacity of PC in GB (hint: In case PC not provided put zero) ≥ 4
Processor speed in GHz (PC) ≥ 3
Frame Grabber incorporated(PC) No
Connectivity between USG and PC No
PC with all software inclusive interfaced with USG machine No
Laser Color printer provided No
Inkjet/Thermal printer provided No
CD/DVD/USB Produced should be playable on any system No
Glossy Colour Print Paper supplied NA
Dicom Interfacability Dicom to PACS Yes
Interfacability Dicom to RIS Yes
Power Requirements Resettable over current breaker shall be fitted for protection Yes
Power backup UPS
Back up time in minutes (minimum) 30.0
Suitable Servo Controlled Stabilizer / CVT should be provided No
Warranty and Maintenance Warranty in Years (Option of comprehensive warranty is available through bidding only, which if opted will supersede normal warranty in the catalogue) 5 Or higher