Catalogue values for SPECTROPHOTOMETER

Category Specification Bid Requirement (Allowed Values)
LIGHT SOURCE Light Source Xenon Flash
Diffraction Grating (lines/mm) 1000
Switching of Light Source Automatic
Wavelength Setting And Scanning Automatic
TYPE Type of Spectrophotometer UV - VISIBLE
Type of Optical System Single Beam
Monochromator Type Czerny-turner
Detector Type CMOS
Spectral Bandwidth Type Fixed Bandwidth
OPTICAL PERFORMANCE Source Wavelength - Minimum 200.0
Source Wavelength - Maximum 320, 340, 760, 780, 900, 960, 1000, 1100, 1500, 2000, 2500, 2700, 3000, 3200, 3300, 1650, 644 Or higher
Resolution 1.0
Wavelength Accuracy (+/-) 1.0
Bandwidth (nm) 1
Focal length (mm) 200 to 249
Photometric measurement modes Concentration (Wavelength, time)
Photometric Absorbance (Max) (Abs) (+/-) 8
Photometric Absorbance Accuracy (Max) (+/-) 0.001 - 0.03 Or lower
Photometric Transmittance (Max) (+/-) 100.0
Photometric Transmittance Accuracy (Max) (+/-) 0.0 - 0.5 Or lower
Photometric Reflectance (Max) (+/-) 0.0
GENERIC Display In-built display
Display type TFT
Display Size 7.0 - 30.0 Or higher
Test result printing Through external printer
If inbuilt printer, Type of printer N/A
Internal storage(GB) 80.0 - 80.0 Or higher
Mximum Path Length of sample 0.2
Number of Cuvettes supplied-inclusive in the scope of supply 0
Spares and Consumables inclusive in the scope of supply No
OPERATING CONDITIONS Minimum Operating Temperature 5.0
Maximum Operating Temperature 50.0
Operating Humidity (RH) (%) at 40 degree C 85.0
WARRANTY Onsite Warranty 5.0
Availability of Spares (at extra cost) 7.0
Certification Availability of UL/CE Certification as per EN 61010-1:2010 Safety requirements for electrical equipment for measurement, control, and laboratory use General requirements No
Conformance to EMC/EMI as per EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use No
Availability of Test Reports from Central Govt / NABL approved / ILAC acredited lab to prove conformity to the specification No